Command Reference
R&S
®
ZNA
1327
User Manual 1178.6462.02 ─ 12
BEFore
is appropriate if the device needs to be triggered before the data
is acquired (typical example: a power meter which is used for
power calibration)
AFTer
is appropriate if the device needs to be triggered just after data
acquisition (typical example: an external generator which must
be reconfigured to get ready for the next measurement)
*RST:
AFTer
Example:
See
TRIGger:CHANnel<Ch>:AUXiliary<n>:STYPe
Options:
R&S
ZNA-B91
Manual operation:
See
TRIGger:CHANnel<Ch>:AUXiliary<n>:STYPe
<Type>
Defines how the selected sweep sequence is indicated at the related Trigger Out con-
nector.
Use
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval
to select the sweep
sequence.
Suffix:
<Ch>
.
Channel number
<n>
1 = Trigger Out A, ..., 4 = Trigger Out D
Parameters:
<Type>
POSitive | NEGative | AHIGh | ALOW
POSitive | NEGative
pulses with positive or negative polarity
AHIGh | ALOW
High (low) voltage level in active state and low (high) level other-
wise
Note
that pulse generator pulses cannot be indicated using high
and low voltage levels.
Example:
See
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval
Options:
R&S
ZNA-B91
Manual operation:
See
8.3.20
New commands for R&S
ZNA
Mixer Phase and Delay Measurements
.............................................................. 1328
.......................................................................................1336
...........................................................................1337
SCPI Command Reference