Trigger Out connectors A to D are provided by the optional
. If this hard-
ware option is not installed, it is not possible to generate outgoing trigger signals. In
particular, the "Trigger Out Manager" dialog is not available in this case.
Sequence
Selects the sequence to be indicated at the related Trigger Out "Connector". Trigger
out signals can be generated for every sweep, segment, partial measurement, and
sweep point.
Furthermore, trigger out signals can be generated for the data acquisition phases and
for the start of pulse generator pulses.
Remote command:
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval
Signal Type
Defines how the selected
is indicated at the related Trigger Out "Connector":
●
using pulses with positive or negative polarity ("Pulse Positive", "Pulse Negative")
●
sending the high voltage level in active state and the low level otherwise ("Active
High"), or vice versa ("Active Low").
Note
that pulse generator pulses cannot be indicated using high and low voltage lev-
els.
Remote command:
TRIGger:CHANnel<Ch>:AUXiliary<n>:STYPe
Delay
Specifies the delay of the trigger signal relative to the selected
If pulses are used as trigger out signals, the delay is defined relative to the start or end
of the sweep event (see
Remote command:
TRIGger:CHANnel<Ch>:AUXiliary<n>:DELay
Width
If pulses are used as trigger out signals (see
on page 435), this parame-
ter specifies the pulse width.
Sweep Softtool