Command Reference
R&S
®
ZNA
1039
User Manual 1178.6462.02 ─ 12
<Format>
COMPlex | LINPhase | LOGPhase
COMPlex - complex values (real and imaginary part).
LINPhase - Linear magnitude and phase.
LOGPhase - dB-magnitude and phase.
<ModeImpedance>
CIMPedance | PIMPedance
CIMPedance
: normalize to the common target impedance (from
options line); this is the default if the parameter is ommitted
PIMPedance
: normalize to the individual port reference impe-
dances
see
"Renormalization of S-parameters"
<Port>, <Port>, ...
Physical port numbers
Example:
Suppose that a full two-port calibration for ports 1 and 2 and
channel 1 has been performed, and that a DUT with two bal-
anced ports is connected. The analyzer measures an arbitrary
mixed mode S-parameter.
MMEM:STOR:TRAC:PORT 1, 'Test_CIMP.s2p',
COMPlex, CIMPedance, 1, 2
Calculate all single-ended S-parameters, renormalize them to
the common target impedance and store them to a two-port
Touchstone file.
MMEM:STOR:TRAC:PORT 1, 'Test_PIMP.s2p',
COMPlex, PIMPedance, 1, 2
Calculate all single-ended S-parameters, renormalize them to to
the individual port reference impedances and store them to a
two-port Touchstone file.
Usage:
Setting only
Manual operation:
See
8.3.12
OUTPut Commands
The
OUTPut...
commands control the characteristics of the analyzer’s output ports.
OUTPut<Ch>:UPORt:SEGMent<Seg>:STATe
.................................................................1041
OUTPut<Ch>:UPORt:SEGMent<Seg>[:VALue]
...............................................................1041
....................................................................................... 1043
OUTPut:ULED:STATe
<ledState>
Configures the "User Defined" LED at the front panel of the R&S
ZNA. This is a persis-
tent setting.
Parameters:
<ledState>
OFF | GREen | RED | FGReen | FRED
SCPI Command Reference