GE Analytical Instruments ©2016
102 of 270
DLM 68100-09 EN Rev. A
Chapter 4: Installation
Specifications —
IP56 Accessory
Typical supply:
Pressure
60 psi (4.9 Bar)
Demand (Flow)
0.36 scfm (10 L/min)
Maximum Pressure
100 psi (7 Bar)
Step 11: Turn on Power and Configure Basic Settings
Enable power to the Analyzer. Before using the Analyzer, you may need to configure various basic settings.
1. Turn on the power switch located on the side of the Analyzer. The display will light up. (It can take up to
two minutes for the
System Info
screen to appear.) If this does not occur, refer to
2. The
System Info
screen appears with a reminder to attach the waste lines. If the waste lines were not
attached already, do that now (see Installation Step 4). When ready, press the
Main
button on the
display.
Setting the Clock
Set the Analyzer clock to ensure that all TOC measurements reflect the correct date and time, as follows.
1. Press the
Menu
button.
2. Select the
Maintenance
tab.
3. Press the
Advanced
button
4. Press the
Clock
button.
5. Specify the correct date and time via each of the buttons.
6. Press the
Save
button, and then press the
Back
button twice.
Step 12: Perform a Test Analysis
To confirm that the Analyzer is in proper working condition, perform a test analysis on a sample for which you
know the approximate TOC or NPOC value and which has a concentration of less than 1,000 ppm.
1. Insert the check standard tube into the sample.
2. Press the
Setup
tab, and then press the
Check Standard
button. Press the
Check Standard Setup
button.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...