GE Analytical Instruments ©2016
140 of 270
DLM 68100-09 EN Rev. A
Chapter 6: Analyzer Operation
3. Touch the protocol name on the screen.
4. To modify this protocol, press the
Modify
button. The
Grab (Setup)
screen appears.
Figure 51: The Grab (Setup) Screen
5. Make changes to the protocol, as needed:
• Press the
Measure
button to specify which value to measure: IC, NPOC, TOC, or TC.
• Press the
Range
button to specify one of the five range buttons, or select the
Auto
button for the
Auto Range function (see
).
• Press the
Acid
and
Oxidizer
buttons to specify the percent of each reagent to use in the protocol.
• Press the
Reps
button to change the default setting for the number of measurements (repetitions)
the Analyzer will make for this sample. If
Repeat Criteria
is
ON
, then this field will be grayed out and
uneditable.
• Press the
Rejects
button to change the default setting for the number of measurements the
Analyzer should exclude from its calculations. If
Repeat Criteria
is
ON
, then this field will be grayed
out and uneditable.
• Press
Repeat Criteria
to allow the Analyzer to follow configurable logic to determine the number of
repetitions to run. If
ON
is selected, you will be able set the maximum number and minimum number
of repetitions for the vial. The number of reps will be determined by the InnovOx depending on what
you enter for the RSD criteria. If you enter 7 Max Reps and 2 Min Reps, the Analyzer will run at least
two repetitions and compare the calculated RSD to the RSD criteria entered. If the calculated RSD is
N
OTE
:
The
number of rejects must be less than the number of repetitions.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...