Chapter 5: Calibration and Verification
7. Enter a name for the new calibration protocol, and then press the
Save
button.
These ranges are based on a completely filled 40-mL Vial and no sample flushes. The maximum
number of repetitions that can be selected is 20. Divide the number of reps shown above by 2 for TOC
and TC modes.
Table 9: Approximate Number of Calibration Repetitions Possible
Calibrating the Analyzer
To calibrate the Analyzer
1. Select the
Cal/Ver
tab.
2. Press the
Calibrate
button.
3. Select a calibration protocol from the list by touching the screen. If you want to create a new
calibration, follow the previous steps in
“To set up a new calibration” on page 116
4. Press the
Start Cal
button and follow the on-screen prompts. The prompts will vary depending on the
settings you have specified in the selected calibration.
5. After the Analyzer has completed measurements on all standards vials, the
Calibration Summary
screen
displays the following values. (Refer to
• Expected - Indicates the expected concentration of the calibration point.
When Auto Dilution is Off, the expected concentration is the Certified Concentration of the standard.
For reagent water, there is no expected concentration, so the entry is “N/A.”
When Auto Dilution is
On
, the Expected concentration is the concentration of the standard after
dilution. For example, if a 1,000 ppm TOC standard is used in the calibration, and is diluted to
calibrate the Analyzer at 100 ppm, the Expected concentration will be 100 ppm.
• Measured: Indicates the average measured concentration of the standard, using default calibration
constants.
• RSD: Indicates the percent relative standard deviation of the non-rejected measurements of the
standard.
• Adjusted: Indicates the calculated concentration of the standard, using the new calibration
constants from the calibration just completed.
Mode
Range
100 ppm
1,000 ppm
5,000 ppm
20,000 ppm
50,000 ppm
NPOC
6
7
14
20
20
TOC
3
3
12
20
20
IC
6
7
14
20
20
TC
3
3
12
20
20
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...