GE Analytical Instruments ©2016
167 of 270
DLM 68100-09 EN Rev. A
Chapter 6: Analyzer Operation
Setting the Clock
Set the Analyzer clock to ensure that all TOC measurements reflect the correct date and time.
To set the clock
1. Select the
Maintenance
tab.
2. Press the
Advanced
button. The
Advanced
screen appears.
3. Press the
Clock
button, and using the keypad that appears specify the correct date and time.
4. Press the
Save
button.
Upgrade Option
You will use the
Upgrade
button on the
Advanced
screen when you install firmware upgrades. For complete
instructions, please refer to the instruction sheet shipped with the firmware USB key.
If you downloaded the upgrade files from the GE website, make sure that the files are saved in the proper folders
on your USB key (:\InnovOx\Upgrade). After you install your USB key in the USB port, wait 15 seconds and then
select the
Upgrade
button. This process may take several minutes.
Gas Valve PID Parameters
The InnovOx Analyzer contains Gas Valve parameters that have been previously set at the factory using the
default setting of PID Set 2. In the case where an instrument is upgraded in the field from firmware versions
≤
3.06 to versions
≥
3.07, the Gas Valve PID parameter may need to be changed. The determining factor is the
style of gas manifold installed in the instrument. Using
, compare the gas manifold in your Analyzer
and set the corresponding Gas Valve PID setting using the following instructions.
To configure gas flow PID option settings
1. Using
, identify the type of gas manifold installed in the Analyzer.
2. Select the
Maintenance
tab.
3. Press the
Advanced
button. The
Advanced
screen appears.
4. Locate and verify which PID option is currently active by identifying whether “PID Set 1” or “PID Set 2”
appears on the
Gas Valve PID
button.
5. To change the PID setting, press the
Gas Valve PID
button. The
PID Set 1
and
PID Set 2
buttons
appear.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...