GE Analytical Instruments ©2016
196 of 270
DLM 68100-09 EN Rev. A
Chapter 8: Maintenance
Table 15: 2-Year Maintenance Worksheet — Sievers InnovOx Online TOC Analyzer
Item
Operating Time
Installation Date
Notes
GLS (clean and refill with DI water)
1 month
Oxidizer reagent
1 month, or as needed
Dilution water
1 month
High pressure valve seals
1 month
High pressure valve seals
2 months
Pinch valve tubing
3 months
High pressure valve seals
3 months
Exterior sample pump tubing (optional
equipment)
3 months
Syringe lubrication
3 months
Reactor peristaltic pump tubing
3 months
Carrier Gas CO
2
Scrubber cartridge (optional
equipment)
3 months
High pressure valve seals
4 months
High pressure valve seals
5 months
Acid reagent
6 months, or as needed
High pressure valve seals
6 months
Exterior sample pump tubing (optional
equipment)
6 months
Syringe lubrication
6 months
Reactor peristaltic pump tubing
6 months
Sample/reagent peristaltic pump tubing
6 months
Waste tubing
6 months
Carrier Gas CO
2
Scrubber cartridge (optional
equipment)
6 months
High pressure valve seals
7 months
High pressure valve seals
8 months
High pressure valve seals
9 months
Exterior sample pump tubing (optional
equipment)
9 months
Syringe lubrication
9 months
Carrier Gas CO
2
Scrubber cartridge
(optional equipment)
9 months
High pressure valve seals
10 months
High pressure valve seals
11 months
High pressure valve seals
12 months
Pinch valve tubing
12 months
Exterior sample pump tubing (optional
equipment)
12 months
NDIR Filter
12 months
Syringe lubrication
12 months
Reactor peristaltic pump tubing
12 months
Sample/reagent peristaltic pump tubing
12 months
Norprene Chemical waste tubing
12 months
Carrier Gas CO
2
Scrubber cartridge
(optional equipment)
12 months
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...