Chapter 7. Operation with the Filter Panel
The Filter Panel is an optional accessory that should be used to filter sample streams that contain high levels of
suspended solids or solids having diameters larger than 200
m. The InnovOx On-Line TOC Analyzer is designed
to control up to five Filter Panels.
Refer to the
GE Filter Panel Operation and Maintenance Manual
for instructions on the installation and setup of
the Filter Panel.
Operation of the Filter Panel with a Sample Pump
The Filter Panel must be installed so that the water to be filtered flows through the Filter Panel. In some cases,
the water pressure in the feed line is not sufficient to flow through the Filter Panel, through the filter screen, and
to the Analyzer. In those cases, it is necessary to pump the filtered water from the Filter Panel to the Analyzer.
The InnovOx On-Line TOC Analyzer can accommodate an optional external peristaltic sample pump to
overcome inadequate line pressure. The pump is controlled by the Analyzer to draw filtered water from the Filter
Panel prior to each Sample Flush and measurement.
Control of Filter Panel
The Analyzer has two parameters related to the operation of the Filter Panel:
Blowback Pulses
and
Sample Flow
Time
. These parameters operate the blowback feature, which automatically cleans the filter screen in the Filter
illustrates a measurement using a Filter Panel. The measurement was set up to include one
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...