Appendix B: Acronyms and Definitions
Acid %
The amount of acid added to the sample, oxidizer, and dilution water (if any). A setting of
1% Acid means that 1% (V/V) of the total solution in the sparger will be acid.
Auto Dilution
The
Auto Dilution
feature is used only during calibrations and verifications. The InnovOx
TOC Analyzer automatically dilutes a standard, having a higher concentration, so that
the Analyzer can be calibrated or verified at a lower concentration.
The standard can be diluted by any factor, up to a limit that differs according to which
measurement Range is to be used:
For example, a standard having a concentration of 1,000 ppm TOC can be diluted down
to ~7 ppm in the 100 ppm range or 100 ppm in the 1,000 ppm range. It can be diluted to
1250 ppm in the 5,000 ppm range. In the 20,000 ppm range or 50,000 ppm Range,
Auto
Dilution
is not active.
Auto Range
The
Auto Range
feature allows samples to be measured when the correct
measurement Range is not known. The InnovOx TOC Analyzer first makes a preliminary
measurement of the sample using the 50,000 ppm Range. Based on the results of that
measurement, the Analyzer selects the correct Range to use and performs the analysis
of the sample using the protocol settings selected by the user. For
Auto Range
to be
applied, all five Ranges must be calibrated, and a calibration for each Range must be
assigned to a Calibration Group (defined later in this section). The Auto range can use
default calibrations.
Auto Restart
The
Auto Restart
feature sets the Analyzer to resume analysis after a power outage.
When the
Auto Restart
feature
is
Off
, the Analyzer will remain in Standby when power is
restored.
Blank Blank
refers to the measurement of the carbon concentrations in the acid, oxidizer, and
dilution water. The Analyzer performs this measurement automatically when the
Blank
button is pressed. The
Blank
button is found on the
Maintenance
screen.
It refers to the calculated contribution that the carbon in the acid, oxidizer, and dilution
water make to a measurement performed by the Analyzer. When
Blank Correction
is
On
during the analysis of a sample or standard, this blank contribution by the reagents and
dilution water is subtracted from the carbon measurement.
The advantage of using the Blank Correction feature is that the correction is calculated,
regardless of any changes in the measurement Range, or changes in Acid or Oxidizer
Range
Maximum Dilution Ratio
Up to 100 ppm
14:1
Up to 1,000 ppm
10:1
Up to 5,000 ppm
4:1
Up to 20,000 ppm
None
Up to 50,000 ppm
None
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...