GE Analytical Instruments ©2016
223 of 270
DLM 68100-09 EN Rev. A
Chapter 9: Troubleshooting
NDIR
Use the
NDIR
screen to test NDIR temperatures to help facilitate troubleshooting.
To test NDIR temperatures
1. On the
Diagnostics
screen, press the
NDIR
button to display the
NDIR
screen with the current source
and detector temperatures (
Figure 97: The Temperatures Display on the NDIR Screen
2. Compare the temperatures appearing on the screen with the following criteria:
• The Source Temperature is maintained at 65
o
C
+
2
o
C.
• The Detector Temperature is maintained at 55
o
C
+
2
o
C.
Heater
The heater test is an examination of the overall reactor heater thermal control system.
To test the heater temperatures
1. On the
Diagnostics
screen, press the
Heater
button to display the
Heater
screen. The
Heater
button is
automatically set to
Off
(gray color) to indicate that the heater is Off. The reactor fan is
On
.
2. To start the test, press the
Heater
button (
). The button turns orange, the heater powers on,
and the system begins going through its programmed heating and cooling profile.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...