Chapter 9: Troubleshooting
•
Check that the sample coil is full of liquid. If excessive bubbles are present, a dilution water flush should
be performed. Watch the operation of the syringe pump to ensure that the syringe is not leaking, and
few if any, bubbles are introduced into the coil.
•
Inspect the syringe pump for leaks, and to confirm that the plunger moves during analysis.
•
Confirm that the peristaltic pumps in the Sample Handling Module and Reactor Module operate during
an analysis.
•
Confirm that the Gas-Liquid Separator contains liquid to the proper level (
) and that all tubes
are properly connected.
•
If maintenance work has recently been performed on the Analyzer, confirm that electronics cables are
securely attached.
Figure 89: Connections to the Gas-Liquid Separator
Reviewing Solutions for Basic Problems
If a visual inspection of the Analyzer does not help you identify the source of the problem you are experiencing,
consult the specific problems discussed in this section.
The Analyzer Will Not Power On
If the power switch is on, but the display is not illuminated, touch the display. The Analyzer may be in Standby to
save power.
Sparger Outlet
Reactor Outlet
Gas-Liquid
Separator (GLS)
Waste
NDIR Outlet
NDIR Filter
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...