Chapter 4: Installation
lines when the Flush is complete. If not, press
Flush All
again. (If one or more of the reagent, dilution
water, or check standard lines are not flushed, see
)
11. When the flush is completed, press the
Back
button to return to the
Menu
screen.
12. Press the
Back
button.
13. When ready, press the
Start Analysis
button. Press
Main
to return to the
Main
screen. A graph of the
CO
2
peaks will be displayed. Each peak appears after its peak height is measured, and the measured
NPOC concentration appears at the top of the screen.
14. When the analysis is complete, the CO
2
peaks will be displayed on the screen. If there are no peaks and
the Mass Response axis is only 1.0 µg full scale, then refer to the Troubleshooting section of this manual.
Otherwise, proceed to the next step.
Step 13: Calibrate the Analyzer
Before placing the Analyzer into normal operation, calibrate the Analyzer in a manner suited to your application.
Consult
Chapter 5, "Calibration and Verification"
for details on calibration options.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...