Chapter 9: Troubleshooting
6431 – Firmware Update Error
Message: “Please shut down the Analyzer and power cycle. Retry the upgrade. If the problem persists, please
export the errors and contact GE Analytical Instruments for further instructions.”
Explanation: This error would occur only when a new version of firmware is being installed in the Analyzer.
6432 – Datastore Error
Message: “Please shut down the Analyzer and power cycle. If the problem persists, please export the errors and
contact GE Analytical Instruments for further instructions.”
Explanation: An error has been found during writing to the data history portion of the flash memory card on the
Analyzer’s System Controller Board. Archiving the
Datastore
manually may alleviate the issue.
for instructions. If this error persists, please contact GE Analytical Instruments.
6433 – Sparger Gas Flow Error
Message: “Go to the
Diagnostics
screen and run the Gas Valves test (V7-SH). If the gas flow rate fails to maintain
200 cc/min, please export the errors and contact GE Analytical Instruments for further instructions.”
Explanation: The flow rate of gas through the sparger was found to be too low. There may be a problem with
valve V7-SH or with the gas tubing. Check that the correct gas flow is exiting the Gas Supply Module.
6434 – Carrier Bypass Gas Flow Error
Message: “Go to the
Diagnostics
screen and run the Gas Valves test (V1-R). If the gas flow rate fails to maintain
165 cc/min, please export the errors and contact GE Analytical Instruments for further instructions.”
Explanation: The flow rate of carrier gas through valve V1-R was found to be too low. There may be a problem
with valve V1-R or with the gas tubing. The NDIR filter also may be plugged. Check that the correct gas flow is
exiting the Gas Supply Module. Valves V4-R or V5-R not opening could also cause this error to appear.
6435 – Reactor Gas Flow Error
Message: “Go to the
Diagnostics
screen and run the Gas Valves test (V2-R). If the gas flow rate fails to maintain
165 cc/min, please export the errors and contact GE Analytical Instruments for further instructions.”
Explanation: The flow rate of carrier gas through valve V2-R was found to be too low. The Reactor may be
plugged. There also may be a problem with valve V2-R, V7-R, or with the gas tubing. The NDIR filter may be
plugged. Check that the correct gas flow is exiting the Gas Supply Module.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...