GE Analytical Instruments ©2016
147 of 270
DLM 68100-09 EN Rev. A
Chapter 6: Analyzer Operation
Figure 52: The Maintenance Tab
1. When the
User Management
feature is enabled, first log in to the InnovOx with the Administrator ID and
password. Otherwise, if the User Management feature is disabled, continue to the next step.
2. Select the
Maintenance
tab and press the
Conversion Factor
button (
) to access the five
available conversion factors (
Factor 1
,
Factor 2
,
Factor 3
,
Factor 4
, and
Factor 5
).
3. Press the
Factor 1
button (or other
Factor
button to configure that factor). The Factor screen appears
with configurable
Factor Equation
,
Factor Units
and
Factor Name
buttons (
).
Figure 53: Configuring a Conversion Factor
4. To set up a Conversion Factor equation, press the
Factor Equation
button and use the keypad to enter
the equation. Use the letter "C" in the equation to represent the raw value produced by the Analyzer. (It
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...