Chapter 4: Installation
Attach a short piece of the same tubing to the Luer-Lok adapter fitting on the pump outlet on the bottom of the
pump head. The other end of the tube is attached to the selected Sample Cup inlet.
Installing the Wastewater Sampler
Figure 14: Wastewater Sampler
The Wastewater Sampler pipe assembly, inlet tubing, and two brackets for the Wastewater Sampler are packed
separately and placed within the Analyzer’s main shipping container. You will need to attach the bracket to the
side of the Analyzer and install the sampler pipe assembly using one of two configurations (velocity method or
the overflow method), as described in the following procedures.
N
OTE
:
In 5-stream Analyzers, Pump #1 may be used to pump Stream #1 and #2. Pump #2 may be used to pump
Stream #3 and #4. Do not mix the Streams and Pumps.
2-Stream Option
5-Stream Option
Original Clear New Opaque
Pipes (left) Pipes (right)
Max Sample Temp Max Sample Temp
60° C (140° F) 85° C (185° F)
Note: In applications with sample temperatures above
85
° C (
185
°
F), the 1-in. flexible sample tubing should
not be used. The samples should be plumbed with rigid
piping suitable for the sample temperatures and any
corrosive properties the sample might exhibit.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...