GE Analytical Instruments ©2016
84 of 270
DLM 68100-09 EN Rev. A
Chapter 4: Installation
Installing the 4-20 mA Connection
The data from the Analyzer may be recorded using the 4-20 mA outputs located on TB1. Each analog output is
assigned to a particular sample stream. 4-20 mA output #1 provides an analog signal proportional to the
concentration measured in sample stream #1. Output #2 provides a signal proportional to the concentration
measured in sample stream #2, and so forth. The analog output is calibrated at the factory prior to shipping. An
example wiring connection for one of the 4-20 mA outputs is shown in
.
Figure 7: Wiring Diagram for the 4-20 mA Connection
Installing the Remote Start (Binary Input) Connection
The Analyzer remote start (binary input) connection, located on TB2, can be wired in a variety of ways and is
intended to execute a start/stop command to the Analyzer from a remote location, such as a Process Logic
Control (PLC) System or Center. Wiring examples are shown in
.
Binary Input using Analyzer’s Internal Supply
shows how the internal 24 VDC of the Analyzer (on TB4) may be
used to power the binary input.
No menu configuration is necessary. For more information on diagnostics testing of the binary input, see
Remote start/stop using binary input works in tandem with the display
Start and Stop Analysis
button.
Changing the state of the
Remote Input
will change the
Analysis Run
state if not already in that mode. The
Display Start/Stop Analysis
button remains functional while the remote start/stop is in use. This allows for
manual override of the analysis mode.
N
OTE
:
For additional information on setting up 4-20 mA outputs in the Analyzer firmware, see
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...