Chapter 8: Maintenance
2. Press the
Flush Reagents
button.
3. Select the flush you want to perform:
•
Flush Acid
: Flushes the acid.
•
Flush Oxid(izer)
: Flushes the oxidizer.
•
Flush Dilution (Water)
: Flushes the dilution water.
•
Flush Check Standard
: Flushes the check standard.
•
Flush All
: Flushes the acid, oxidizer, check standard, and dilution water.
Cleaning the Analyzer
You can clean the external housing of the Analyzer with a damp cloth using water or non-abrasive cleaners.
Turn off power to the Analyzer and disconnect it from main power prior to cleaning. Do not spray liquids directly
on the Analyzer. Wipe dry with a clean, soft cloth.
The touchscreen may be cleaned with a soft cloth or cotton pad lightly applied with a spray cleaner. Isopropyl
Alcohol may be used, but ensure that all solvent residue is removed. Water may cause damage or discoloration
of the polarizer. Clean any condensation or moisture from any source immediately. Before cleaning the screen,
you can temporarily deactivate touchscreen sensitivity for 30 seconds.
To deactivate the screen
1. Press the
Menu
button and select the
Maintenance
tab.
2. Press the
Advanced
button, and then press the
Clean Screen
button.
Cleaning the Gas-Liquid Separator
The gas-liquid separator requires periodic cleaning, when residue builds up inside. Water in the gas-liquid
separator is acidic. Refer to
Figure 73: The Gas-Liquid Separator on page 182
, as needed.
To clean the gas-liquid separator
1. Shut down the Analyzer via the on-screen menu (
Maintenance
Shutdown
).
N
OTE
:
The Check Standard must be the
Enabled
state.
Warning
Contents are acidic. Wear eye protection and gloves during this procedure.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...