GE Analytical Instruments ©2016
175 of 270
DLM 68100-09 EN Rev. A
Chapter 7: Operation with the Filter Panel
4. Confirm that the pneumatic-signal air tube is connected to the Filter Panel and to the appropriate port
on the left side of the Analyzer, marked
Air to Filter Panel
. If the Filter Panel is filtering Stream 1, the
pneumatic-signal air tube should be inserted in
Air to Filter Panel
port 1.
5. Confirm that compressed air is connected to the
Blowback Air
port on the Filter Panel.
6. Go to the
Stream
(setup) screen (
Setup
tab >
Online
>
Configure Online
) for the stream where the Filter
Panel is installed (
).
Figure 70: Stream (Setup) Screen
7. Press the
Flush/Blowback Setup
button. The
Flush/Blowback Setup
screen appears (
).
Figure 71: Flush/Blowback Screen
8. Press the
Dilution Water Flush
button and then press either the
On
or
Off
button that appears.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...