GE Analytical Instruments ©2016
206 of 270
DLM 68100-09 EN Rev. A
Chapter 9: Troubleshooting
that all connections on gray ribbon cable are secured to their respective boards. If the problem persists, please
export the errors and contact GEAI for further instructions.”
Explanation: The bus referred to in this discussion is the communications bus internal to the Analyzer. This error
will also occur if an AutoRange protocol is started and one or more calibrations in the Calibration group have
been deleted. Archiving the
Datastore
manually may alleviate the issue.
See “Archiving the Data History” on
for instructions. If this error persists, please contact GE Analytical Instruments.
6428 – Flash Card Error
Message: “An error has been detected within the flash card. Shut down the Analyzer and power cycle. To avoid
re-occurrence, always shut down the Analyzer prior to turning off the power. If the problem persists, please
export the errors and contact GE Analytical Instruments for further instructions.”
Explanation: An error has been detected in the flash memory card on the Analyzer’s System Controller Board.
6429 – Network Error
Message: “This is an error that occurs when the InnovOx is trying to communicate with a network. Shut down
the Analyzer, turn off the power and reconnect the network cable. You may need to contact your IT department
for more information. If the problem persists, please export the errors and contact GE Analytical Instruments for
further instructions.”
Explanation: An error inside the Analyzer has been detected that prevents communications with a network.
6430 – Touchscreen Error
Message: “Please shut down the Analyzer and power cycle. If the problem persists, shut down the Analyzer and
power down. Then, check the cable connections. You may be able to regain control of the unit by plugging a
mouse into any of the USB port. If the problem persists please export the errors and contact GE Analytical
Instruments for further instructions.”
Explanation: This error can be troubleshot by shutting down the Analyzer and turning off the power. Access the
display assembly, and inspect the connections and the gasket that surrounds the touchscreen. Re-seat the
connections and realign the gasket. If the problem persists, contact GE Analytical Instruments.
N
OTE
:
It is not abnormal for the yellow LED on the NDIR board to blink faster. However, on Analyzer start-up the
rate should be consistent with the other boards’ LEDs.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...