GE Analytical Instruments ©2016
259 of 270
DLM 68100-09 EN Rev. A
Appendix D: Modbus Map
Register Type
Description
Function
Address
Notes
Discrete Inputs (read)
Status
2
1000
New Data
Bit 0
Error
Bit 1
Standby
Bit 2
Alarm 1
Bit 3
Alarm 2
Bit 4
Alarm 3
Bit 5
Alarm 4
Bit 6
Coils (read)
State
1
2000
See Footnote 1
Run
Bit 0
Stop
Bit 1
Check Standard Enabled
2002
See Footnote 2
Stream 1 Enabled
2003
Stream 2 Enabled
2004
Stream 3 Enabled
2005
Stream 4 Enabled
2006
Stream 5 Enabled
2007
Reject Measurement
2008
Coils (write)
Run
5
2000
See Footnote 3
Stop
2001
Check Standard Enabled
2002
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...