GE Analytical Instruments ©2016
91 of 270
DLM 68100-09 EN Rev. A
Chapter 4: Installation
To install the Wastewater Sampler (velocity method)
Figure 15: Velocity Method
Use this method when the sample flow-through rate is in the range of 2 to 6 GPM (0.454 to 1.36 m
3
/hour) with
minimal back pressure to ensure a high velocity, but with relative laminar flow.
The sample stream to the Wastewater Sampler is installed at the pipe farthest from the instrument. The water
quickly flows over the weir and sample is drawn from the stream in the opposite direction of the sample flow,
allowing solids to flow by the sample tube and preventing them from being drawn into the instrument. See
Flexible 1” tubing is included with the Kit for the supply and drain side of the Wastewater Sampler pipe
assembly. When configuring for velocity mode operation, use the 3/4" NPT to 1" barb fittings for the outlet pipe
of the sampler.
1. Attach and secure the Wastewater sampler brackets to the left side of the Analyzer, as follows. (The
screws are taped to the side of each bracket.)
•
Top bracket
— Slide the following bracket over the plate adapter and secure it by tightening a screw
on each end of the plate adapter.
Figure 16: Top Wastewater Sampler Bracket w/Plate Adapter
Outlet
Sample Flow
Inlet Sample
Flow
Sipper Tube
Pre-drilled screw
holes
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...