Chapter 4: Installation
Table 3: Analog Outputs (TB1)
Pin
Signal
Function
1
Analog 1 (+)
4-20mA Stream #1
2
GND_ISO (-)
3
ANALOG 2 (+)
4-20mA Stream #2
4
GND_ISO (-)
5
ANALOG 3 (+)
4-20mA Stream #3
6
GND_ISO (-)
7
ANALOG 4 (+)
4-20mA Stream #4
8
GND_ISO (-)
9
ANALOG 5 (+)
4-20mA Stream #5
10
GND_ISO (-)
11
ANALOG 6 (+)
Stream ID
(11 mA, 12 mA, 13 mA, 14 mA, or 15 mA, as
12
GND_ISO (-)
N
OTE
:
4-20 mA outputs are isolated from system ground.
4-20 mA outputs are not individually isolated from each other and, therefore, are not intended to be
connected to multiple systems with differences in ground potential.
4-20 mA outputs may be configured by the user to output NPOC, IC, TC, or TOC ppm levels.
Table 4: Stream ID
Active Stream
Stream ID
Standby
Output matches the
standby
level set for the first
sequentially-ordered active stream, starting with Stream 1
1
11 mA
2
12 mA
3
13 mA
4
14 mA
5
15 mA
Error
Output matches the
error
level set for the first
sequentially-ordered active stream, starting with Stream 1
Warning
n/a (Does not change current value.)
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...