GE Analytical Instruments ©2016
112 of 270
DLM 68100-09 EN Rev. A
Chapter 5: Calibration and Verification
Figure 34: Linear Fit Six-Point Calibration
Point-to-Point Multi-Point Calibration
The advantage of selecting the point-to-point calibration is that any non-linearity of response is compensated
for during the calibration.
depicts a point-to-point calibration with 6 points. Like the linear fit, the point having the lowest
concentration is measured first, and the points must increase in concentration at each step. Point 1 can either
be reagent water or a higher, known concentration.
The Analyzer calculates an intercept and slope for the line connecting each pair of points, as shown in
When this calibration is later applied to a sample, the Analyzer applies the appropriate intercept and slope,
based on the magnitude of its mass response for that sample.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...