GE Analytical Instruments ©2016
202 of 270
DLM 68100-09 EN Rev. A
Chapter 9: Troubleshooting
6405 - Peristaltic Pump Life Warning
Message: "The estimated pump life is less than 2 weeks. Order new peristaltic pumps to replace P1-SH and P1-R."
Explanation: Based on the history of usage, the Analyzer estimates that only two weeks of peristaltic pump life
remains. Order two new peristaltic pumps from GE now to avoid down time when the pumps expire.
6406 - Tubing Life Warning
Message: "The estimated tubing life is less than 2 weeks. Order new tubing."
Explanation: Based on the history of usage, the Analyzer estimates that only two weeks of pinch valve tubing
remains. The tubing must be replaced to protect the operation of the pinch valves. Order tubing from GE now to
avoid down time and more costly failures of pinch valves.
6408 - iButton Read Error
Message: "Unable to read system information from iButton
®
. Please power cycle the Analyzer. If the problem
persists, consult the Operation and Maintenance Manual for further instructions."
Explanation: This is a failure to read information from the iButton. This information includes the serial number,
family, and model. Possible reasons for failure include corrosion on iButton contacts, a gap between the iButton
and system board contacts, or a component failure. Consult GE Analytical Instruments for further instructions on
determining the root cause.
6409 - Internal Communications Error
Message: "Internal communications problem. Please power cycle the Analyzer. If the problem persists, consult
the Operation and Maintenance Manual for further instructions."
Explanation: A module in the Analyzer is not communicating over the bus. First, evaluate the LEDs on the
individual boards. You can sometimes view the reflection of the LEDs looking through the hole on the back cover
of the instrument. Or, for a better view, remove the back cover. All boards should have a green LED to indicate
power. Any yellow LEDs blinking faster than once per second, other than on the NDIR board (see figure 55), may
indicate a problem with the board. If this is the case, shut down the Analyzer and turn off power. Check that all
connections on the gray ribbon cable are secured to their respective boards. If the problem persists, export the
errors and contact GE Analytical Instruments for further instructions.
N
OTE
:
It is not abnormal for the yellow LED on the NDIR board to blink faster. However, on
Analyzer start-up the rate should be consistent with the other boards’ LEDs.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...