7. Remove the plug located at the top of the waste drain outlet pipe (the pipe without the internal sipper
tube) and install a filter vent in its place for each stream you are installing.
8. Insert a Wastewater Sampler pipe assembly into the clamp located on the outside of the Analyzer. For
this overflow method, ensure that the waste drain pipe (without internal needle)
is installed farthest
from the instrument
.
Also, ensure that the sample inlet pipe is installed nearest to the instrument.
Figure 27: Overflow Method — Installation
9. Close the two clamps over the pipe and use pliers or channel locks to secure.
10. Adjust the height of the sipper tube to enable drawing sample from a laminar flow without bubbles.
Figure 28: Overflow Method — Adjusting the Sipper Tube Height
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...