GE Analytical Instruments ©2016
53 of 270
DLM 68100-09 EN Rev. A
Chapter 2: System Description
Additional System Specifications
Carrier Gas
CO
2
free air or Nitrogen; Gas must be oil- and water-free;
Delivery Pressure of 50-100 psi required when using the air purifier or
Carrier Gas CO
2
Scrubber options; Delivery pressure must be regulated
to 30-100 psi when using a cylinder of purified gas; Unit will consume
up to 200 cc/min in NPOC mode.
Sample Analysis Volume
Range
Sample Volume
Up to 100 ppm
2.82 mL
Up to 1,000 ppm
2.00 mL
Up to 5,000 ppm
0.80 mL
Up to 20,000 ppm
0.20 mL
Up to 50,000 ppm
0.08 mL
Operating Specifications
Sample Time (min)
NPOC (OL)
TOC & TC (OL)
(Time to reported value)
Rep 1
8.3
9.7
Rep 2+
5.8
8.3
Sample 1 Flush
+ 1.4 per rep
+ 1.4 per rep
Sample 2 Flush
+ 5.2 per rep
+ 5.2 per rep
Dilution Flush
+ 0.7 per rep
+ 0.7 per rep
Note: The time for the analytical cycle is the same whether or not the
Sample Backflush feature is utilized.
Sample Consumption
Ox OL Sample
50 mL/min and
125 psig
(
See aliquot in respective range
)
Filter Panel
>1.0 L/min w/o exterior sample pump
400 mL/min with exterior sample pump
Temperature
60 °C
Fast Flush
10 mL
Sample Flush
Refer to aliquot volumes in respective range.
Monthly Carrier Gas
Supply Pressure
30 to 100 psig
Consumption
200 cc/min in NPOC mode, Less in TOC mode.
Standard cylinder holds 300 cu ft, or 30-days
worth
Note: The consumption rate is not relevant to the compressed air needed
to operate the Regenerative Air Purifier option. In addition, this flow rate
is not sufficient to operate the IP-56 option. GEAI does not recommend
using cylinder air for the IP-56 option.
Electrical
Power
100 - 240 VAC +/-10% 50/60 Hz, 400 VA
Alarm Relays
30 VDC @ 1A max or 125
VAC @
0.5A max
Binary Input
5-24 V (remote start)
Binary Output
Relay rated max 125 VAC or 110 VDC @ 1A
(analysis complete)
Analog Output
4-20 mA, max load @ 600 ohms, powered
Ethernet
Modbus port 502
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...