GE Analytical Instruments ©2016
189 of 270
DLM 68100-09 EN Rev. A
Chapter 8: Maintenance
3. To facilitate removal of the tubing from the valves, navigate to the
Flow Control Valves
screen (
Menu
Maintenance
Diagnostics
Valves
Flow Control Valves
).
4. Press the button for the appropriate flow control valve module you are changing [
Multi Stream Valves
,
Sample Handler Valves
, or
Reactor Valves
] to display the available valves for that module.
5. Press the specific valve button (such as
V1-SH
on the
Sample Handler Valve
detail screen) to open the
valve. (All valves, except V7-R, have a default state of closed so the first press on the button opens the
valve and the next press closes the valve. V7-R is normally open so the first press will close the valve,)
6. Pull the sample tubing out of the valve and then disconnect the tubing from the barb at both ends.
7. Connect each new piece of tubing to the barbs, and then into the valve. Be sure to seat the tubing fully
under the pin and all the way to the back of the valve opening.
8. After connecting each piece of tubing, return to the Analyzer screen to toggle the valve closed again.
9. After replacing the last piece of tubing, press the
Back
button to return to previous screens and reach
the
Maintenance
tab screen.
10. Enter the installation date for the new sample tubing. See
“Entering Consumables Installation
11. Perform other maintenance tasks, or close the Analyzer door and return normal Analyzer operations.
N
OTE
:
identifies the valves and their locations.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...