GE Analytical Instruments ©2016
60 of 270
DLM 68100-09 EN Rev. A
Chapter 2: System Description
Additional System Components
System/Controller and Electronics
Operation of the Analyzer is controlled by an advanced X86 microprocessor on the System/Controller Board,
which communicates with the other Analyzer modules over a Multi-point Low Voltage Differential Signaling
(M-LVDS) bus. Other electronic assemblies provide the interface, monitor and control functions for the Analyzer:
•
Color LCD VGA display
with touch panel
•
Multistream module board
for microprocessor-controlled selection of the sample to be measured
(This board also controls operation of the optional sample pump accessory.)
•
Sample handling module board
for microprocessor-controlled preparation of the sample,
reagents, and dilution water
•
Gas control module board
for microprocessor-controlled control of the carrier and sparge gases
•
Reactor module board
for microprocessor-controlled oxidation of the samples
•
NDIR module board
for microprocessor-controlled measurement of the CO
2
produced in the
Sample Handling and Reactor Modules
•
Input/Output board
for processing digital input signals and analog and digital outputs
•
Terminator board
, which terminates the M-LVDS bus and provides Ethernet and USB interfaces
• Power Entry/Supply
AFLAS
®
O
‐
rings
on
seals
Norprene
Chemical
with
PVC
lining
Waste
tubing
PEEK
®
Sample
manifold,
reagent
inlet
fittings
PVC
Gas
‐
liquid
separator,
NDIR
detector,
sparger,
multistream
manifold,
sample
cups,
external
tubing
PCTFE,
PTFE
Vial
port
valve
Kalrez
®
and
PEEK
Reagent
valves
Table 1: InnovOx Wetted Materials List
Material
Components
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...