GE Analytical Instruments ©2016
132 of 270
DLM 68100-09 EN Rev. A
Chapter 6: Analyzer Operation
When multiple streams are measured, the Analyzer measures the enabled streams in numerical order. For
example, Stream 2 follows Stream 1. Any disabled streams are skipped. If the Fail Safe option is employed,
streams are also skipped whenever a stream stops flowing. When flow resumes in that stream, the Analyzer will
measure it again in its original sequence.
To set up stream measurements
1. On the
Setup
tab, press the
Online
button.
2. Press the
Configure Online
button. The
Configure Online
screen appears.
3. Press a stream button (such as
Stream 1 Setup
) to configure making measurements for this stream.
The setup screen for the selected stream appears.
Figure 46: Stream 1 Screen
4. Press the
Range
button to display the range buttons. Press a button for the specified range (such as
Up
To 1,000
) or press
Auto
to have the Analyzer use the Auto Range function. (For more information on the
Auto Range function, see
.)
5. Press the
Measure
button to specify which value to measure: IC, NPOC, TOC, or TC.
6. Press the
Advanced Setup
button
to display the following options:
•
Blank Correction
button — Set the blank correction feature to
On
or
Off
.
N
OTE
:
To increase the accuracy and speed of the measurements, observe the concentrations in adjacent
sample streams. Minimize the concentration difference between adjacent streams. It takes more
time to get an accurate measurement because more Sample Flushes are required, when one
stream is a much higher concentration than the following stream.
N
OTE
:
If the stream button is disabled, press the
Enabled
button that appears to the right. At least
one stream must always be enabled in Online mode.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...