GE Analytical Instruments ©2016
133 of 270
DLM 68100-09 EN Rev. A
Chapter 6: Analyzer Operation
•
Sparge
button — Change the default sparging time (in minutes), when measuring in NPOC mode.
•
Outlier Setup
button — Set up this feature to ignore any value above a set
outlier
threshold for one
to three consecutive measurements. (The value is NOT stored in the in the instrument nor viewable
on the data screen.) Configure this outlier rejection feature (for use when operating in NPOC mode)
by setting the following:
•
Outlier Rejection
button — Enable or disable the Outlier feature. The default value is
Disabled
. If
you switch modes (to IC, TC, or TOC) and then return to the NPOC, you will need to enable this
feature again.
•
Outlier Threshold
button — Set the value (ppm) that when surpassed will not be stored by the
instrument nor appear on the data screen.
•
Rejected Outliers
button — Set the number of consecutive outliers (1, 2, or 3) acceptable, before
the instrument stores the value and displays it on the data screen.
For additional information on this feature and its applicability, refer to
7. Press the
Acid
button to specify the percent of this reagent to use in the protocol.
8. Press the
Oxidizer
button to specify the percent of this reagent to use in the protocol (NPOC, TOC, or TC
Modes Only). For additional information on what Oxidizer settings to use, see step
and
9. Press the
Reps
button to change the default setting for the number of measurements (repetitions) the
Analyzer will make for this stream.
10. Press the
Flush/Blowback Setup
button to configure flushes and/or the blowback operation of the
optional Filter Panel. (See
Chapter 7, "Operation with the Filter Panel"
for information on the blowback
operation.)
• Press the
Dilution Water Flush
button to rinse out salts or particulates from the reactor and its
seals.
• Press the
Sample Flushes
button to flush the Analyzer with the new sample. Up to 5 Sample Flushes
may be selected. Use Sample Flushes to quickly reach steady-state response when the Analyzer will
be switching from one stream to another and the two streams have very different concentrations.
• If you have a GE InnovOx Filter Panel installed (ONLY), press the
Blowback Pulses
button to set the
number of pulses.
• If you have a GE InnovOx Filter Panel installed (ONLY), press the
Blowback Delay
button to define the
number of seconds for delay between the last pulse and start of the measurement.
11. Press the
Analog Out Setup
button to configure the 4-20 mA analog output for Stream 1. The
Analog
Output Stream 1
screen appears with the following options:
•
Range Max
(ppm)
and
Range Min
(ppm)
buttons — Set the maximum and minimum concentrations
(in ppm) that will cause currents of 20 mA and 4 mA to be output, respectively.
•
Error Level, Warning Level,
and
Standby Level
buttons — Set
the current to be output when the
Analyzer detects an Error, Warning, and when it is in Standby (that is not measuring sample
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...