Chapter 9: Troubleshooting
damage. 4) Thoroughly dry the sensor, the mount, and surrounding surfaces. 5) Reinsert the sensor into its
mount. 6) Using the Analyzer’s touch panel, reset the error (
Maintenance
tab >
Errors/Warnings
button).
6443 - Fluid Leak in Optional C1D2 or ATEX Enclosure Error
Message: "Fluid leak detected in optional outer enclosure. Resolve the source of the leak. Clean and dry the
sensor.
Explanation: The liquid sensor is a corrosion-resistant and optically-based sensor intended for detecting large
leaks inside the optional outer enclosure. The sensor is mounted on the lower-left inside wall of the outer
enclosure and triggers when the face of the sensor contacts fluid. (Due to variations in installation conditions,
fluid from leaks may not immediately contact the sensor.) When a leak is detected, the Analyzer reports this
error, displays a pop-up message, and stops analysis. Do the following: 1) Identify and repair the source of the
leak. 2) Wipe up the fluid inside the outer enclosure. 3) Remove the sensor from its mount, making sure to pull it
straight up to release and protect it from damage. 4) Thoroughly dry the sensor, the mount, and surrounding
surfaces. 5) Reinsert the sensor into its mount. 6) Using the Analyzer’s touch panel, reset the error (
Maintenance
tab >
Errors/Warnings
button).
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...