Chapter 6: Analyzer Operation
To set check standard measurements (Scheduled)
1. On the
Configure Online
screen (
Setup
tab>
Online
button>
Configure Online
button), press the
Check
Standard
button. The
Check Standard
screen appears.
Figure 47: Check Standard Setup Screen
2. Press the
Frequency
button to set how often to measure the check standard.
3. Press the
Start Time
button to set the start time using a 24-hr. clock format (e.g., use 13:00 for 1 p.m.).
4. Press the
Concentration
button to set the ppm concentration of the check standard.
5. Press the
Tolerance
button to set the acceptable range of concentrations for the measurement of the
standard. For example, if the
Tolerance
is set at 20%, then the acceptable concentration is 20% below
or 20% above the
Concentration
value.
6. Press the
Check Std. Setup
button. The following buttons appear on the
Check Standard
screen.
Figure 48: Check Standard (with Setup Buttons) Screen
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...