Chapter 9: Troubleshooting
Pumps
You can test the Analyzer’s pumps using the
Diagnostics
screen. External Pump 1 and 2 refer to the optional
Peristaltic Sample Pumps mounted on the left exterior side of the Analyzer. Each external pump has the capacity
to service two streams.
P1-SH and P1-R are the sample pumps located in the interior of the Analyzer.
To test a pump
1. On the
Diagnostics
screen (
Maintenance
Diagnostics
), press the
Pumps
button. The
Pumps
screen
appears.
Figure 109: The Pumps Screen
2. To test the external pump1, press the
External Pump 1
button. The pump will start pumping. It is
normal for the pump to ramp up its speed when the button is pressed to start the test.
3. Press the button again to stop the pump. The pump will ramp down its speed to stop.
4. Repeat steps
to test each pump installed on the Analyzer.
5. Press the
Back
button to return to the
Diagnostics
screen.
N
OTE
:
The External Pump 1 and External Pump 2 buttons only appear on the screen if your Analyzer has one or
two optional pumps installed. If the Analyzer has only one external pump installed, the button appears as
“External Pump 1.”
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...