GE Analytical Instruments ©2016
141 of 270
DLM 68100-09 EN Rev. A
Chapter 6: Analyzer Operation
less than the RSD criteria, the Analyzer will stop analysis on the vial. If the calculated RSD is higher
than the RSD criteria, the Analyzer will run one more repetition and recalculate the RSD. The number
of reps used to calculated the RSD will be the Min Reps value in consecutive order. If the Analyzer
continues to not satisfy the RSD criteria, it will stop after the Max Reps number is realized and list the
final value in the report.
• Press the
Flush
button if you want to include one to five sample flushes, or a dilution water flush, in
the protocol. This flush occurs prior to the first measurement.
• Press the
Conversion Factor
button to apply a configurable factor to the final result. You can
configure up to five conversion factor equations with unique alphanumeric units by clicking on
Maintenance
tab
Conversion Factors
.
• Press
Advanced Setup
to access
Blank Correction
,
Manual Dilution
, and
Sparge
settings.
Blank
Correction
can be set for
On
or
Off.
Manual Dilution
is used to specify if you have diluted the
sample. The default setting of 1.0 means that the sample was not diluted manually. The final result
will be reported as full strength. The
Sparge
default setting of
0.8 Min
has been determined to be a
sufficient sparge time for most sample matrices (with the exception of the
Up to 100 ppm
range,
which has a default setting of 2 min.) and is only available for NPOC mode.
• Press the
Calibration
button to specify which calibration to use in this protocol.
• Press the
Save
button to overwrite the modified protocol or press the
Save As
button to save as a
unique name.
• Enter a name for the new protocol, and then press the
Save
button. If the
Save As
button is selected,
you will be prompted to enter a new name. The new protocol now appears on the
Protocol
screen.
6. Press the
Start Analysis
button. The
Sample Name
screen with a keypad appears. Enter the name of the
sample for future reference, and press the
Enter
button. The Analyzer will start analysis and the
Setup
screen appears. A “Running Grab” message also appears at the top-left of the
Menu
screen.
7. Press the
Main
button to view the CO
2
peaks or analysis data.
The following table lists the approximate number of repetitions that can be performed with the Analyzer when
the vial port is used to analyze samples in 40-mL vials. It is recommended that you completely fill the vial to
maximize the number of repetitions.
The number of repetitions shown for Auto Range refers to when the sample is less than 1,000 ppm, and the
instrument selects the 1,000 ppm range. If a higher range is selected, more repetitions can be made. The
maximum number of repetitions that can be selected is 20.
N
OTE
:
Only the calibration profiles for the appropriate analysis type and range appear for selection, as
specified via the
Range
button. When the Auto Range function is initiated, you can specify the
Calibration Group
to use with the protocol. For more information, see
N
OTE
:
As an example, the following table lists the approximate number of repetitions
possible when using 40-mL vials. The possible number of repetitions varies according
to the size of the vials used in the protocol.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...