GE Analytical Instruments ©2016
224 of 270
DLM 68100-09 EN Rev. A
Chapter 9: Troubleshooting
The test time is 4 minutes and 50 seconds. At the end of this time frame, the measured temperature
value appears.
Figure 98: The Heater Screen
Heater Testing Criteria
•
Heating Phase — The heating phase lasts 130 seconds, after which the heater reaches its target
temperature of 355° ±5°C. A message indicating the actual heater temperature and results of the test is
shown on the screen.
•
Holding Phase — The heater maintains the set temperature for an additional 50 seconds.
•
Cooling Phase — At 180 seconds, the cooling fan comes on and remains on for 110 seconds. After
which, a message indicating the heater temperature and test results appears, the timer resets to zero,
and the
Heater
button becomes active again.
N
OTE
:
After the test is started, the
Heater
button becomes inactive to allow the thermal control system to
complete its heating and cooling cycle. If the
Back
button is pressed while the heater is heating, the fan
will not cool it down and the heater will be slow to cool. To quickly cool the heater, access the Diagnostics
menu and press the
Heater
button again to activate the fan.
N
OTE
:
The 355° ±5° temperature is an indirect measurement of a controlled reactor temperature. This control
point has been demonstrated to achieve a reactor temperature of 375° C necessary to create
supercritical water in the Analyzer’s sample flow path. For additional information on the Analyzer’s
sample flow path, see
.
Содержание Sievers Innovox
Страница 214: ...GE Analytical Instruments 2016 214 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 88 Analyzer Schematic ...
Страница 235: ...GE Analytical Instruments 2016 235 of 270 DLM 68100 09 EN Rev A Chapter 9 Troubleshooting Figure 104 Reactor Leak Check ...
Страница 248: ...GE Analytical Instruments 2016 248 of 270 DLM 68100 09 EN Rev A Figure 112 Analyzer Dimensions Right Side View ...
Страница 249: ...GE Analytical Instruments 2016 249 of 270 DLM 68100 09 EN Rev A Figure 113 Analyzer Dimensions Left Side View ...
Страница 250: ...GE Analytical Instruments 2016 250 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 258: ...GE Analytical Instruments 2016 258 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 264: ...GE Analytical Instruments 2016 264 of 270 DLM 68100 09 EN Rev A This page is intentionally left blank ...
Страница 268: ...GE Analytical Instruments 2016 268 of 270 DLM 68100 09 EN Rev A Index ...
Страница 269: ...GE Analytical Instruments 2016270 269 of 270 DLM 68100 09 EN Rev A Notes ...
Страница 270: ...GE Analytical Instruments 2016270 270 of 270 DLM 68100 09 EN Rev A Notes ...