Chapter 8 Replacement procedures
8 - 95
8-6-2-5
Monitor removal - R5.x and later (Ergotron Arm)
This procedure applies LOGIQ E9 consoles with Ergortron Arms ONLY. If the LOGIQ E9 has an original
Arm or a Daeil Arm, see:
8-6-2-1 "LCD Monitor removal - R3.x and earlier" on page 8-87
and
"LCD Monitor installation - R3.x and earlier" on page 8-89
, for original Arm,
or
and
8-6-2-4 "LCD Monitor V2 installation - R4.x" on page 8-93
Arm.
A cover at the rear side of the Monitor assembly covers the cable to the monitor. To get access to the
cable connector, remove the Cable Cover.
NOTE:
For R6 and later, the Main Monitor could be a 23 inch Monitor or a 22 inch OLED Monitor. The
Monitor mounts to the Monitor Arm in the same manner as R5, only the type and location of the
cables are different. This procedure covers both R5 and R6 and later.
Table 8-63 LOGIQ E9 Monitor Arms
Original Arm
Daeil
Ergotron