GE
D
IRECTION
5573152-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 8 Replacement procedures
8 - 49
8-5-10
Filter Cover replacement
8-5-10-1
Manpower
One person, 15 minutes.
8-5-10-2
Tools
No tools are needed to replace the filter cover.
8-5-10-3
Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the system
2.) Disconnect the mains power cable from the wall outlet.
Follow this link if you need more information:
4-2-3 "Power shut down" on page 4-6
.
8-5-10-4
Filter Cover removal
1.) Gently pull the Filter Cover out and away from the System.
8-5-10-5
Back Filter Cover installation
1.) Inspect the Filter, and clean if necessary.
2.) Place the Filter Cover into position and press the Filter Cover top corners until the Filter Cover locks
engage.
Figure 8-14 Remove the Filter Cover
Содержание LOGIQ E9
Страница 1: ...12 13 18 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL VERSION R6 Part Number 5573152 100 Revision Rev 2 ...
Страница 2: ......
Страница 11: ...GE DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL xi ZH CN KO ...
Страница 140: ...DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 50 Section 4 8 Site Log This page was intentionally left blank ...
Страница 753: ......
Страница 754: ......