GE
D
IRECTION
5573152-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
8 - 232
Section 8-9 - BEP (Back End Processor) parts replacement
8-9-6-1
Battery Pack - BEP6.x replacement
Table 8-146 Battery Pack - BEP6.x removal
Steps
Corresponding Graphic
1.
Loosen the thumb screws or screws at the top of the BEP cover, or the three screws that attach
the cover to the BEP.
Tilt the top of the BEP cover away from the BEP.
2.
Disconnect the battery from
PCN1
on the
the CB (ChargeBoard).
Position the cable connector into the
opening of the battery cavity, so the cable
and battery can be easily removed.
Close BEP door and secure closed.
BEP6 CB cables
3.
Remove the four screws to the Battery
Cover on the BEP cover.
Remove the Battery Pack Cover.
BEP Cover, BEP Battery Pack Cover screws
4.
Remove the stabilizing foam strip beside the battery pack.
5.
Remove the BEP Battery Pack.