GE
D
IRECTION
5573152-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
8 - 204
Section 8-8 - Casters and Brakes replacement
8-8-3
Front Casters replacement
8-8-3-1
Manpower
One person, 15 minutes.
8-8-3-2
Tools
8-2-5 "Tools needed for servicing the LOGIQ E9" on page 8-5
8-8-3-3
Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the system
2.) Disconnect the mains power cable from the wall outlet.
3.) Disconnect all Probes and External I/O Cabling.
4.) Remove both Side Covers.
5.) Remove the Foot Rest Bumper.
Follow these links if you need more information:
•
4-2-3 "Power shut down" on page 4-6.
•
8-5-3 "Side Covers replacement" on page 8-28.
•