2.
Flip the Fan Tray over on a clean, safe
surface.
Pull out the Filter latch and slide the Filter
out.
Clean the Bottom Filter after it is removed
by removing excess lint or dust from the
soiled side. If necessary, use a clean, soft
brush; or vacuum.
3.
Re-install the Filter into the Fan Tray and re-install the Fan Tray.
Make sure the Fan Tray seats
completely into the Card Cage.
Table 8-123 Removing the Fan Tray Filter