GE
D
IRECTION
5573152-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
10 - 24
Section 10-7 - Electrical Safety Tests
10-7-9-1
Isolated lead (sink) leakage test record
Table 10-15 on page 10-24 shows a typical format for recording the isolated patient lead sink leakage
current.
Measurements should be recorded for the full lead combination under each set of test conditions
specified in:
•
•
as applicable.
Record all data on the Inspection Certificate.
NOTE:
Values in italics font are given as examples only.
1.) Connect Safety analyzer to wall AC power outlet
2.) Plug the equipment under test power cable into the receptacle on the panel of the meter.
3.) Connect the ECG cable to the scanner and the Patient leads to the analyzer
4.) Select the Lead isolation or main applied function on the meter
NOTE:
Consult the manufacturer’s user manual of the Safety Analyzer
•
Test opening and closing the ground with the scanner on and off as indicated in Table 10-9 or
as applicable
NOTE:
Consult the manufacturer’s user manual of the Safety Analyzer
Table 10-15 Typical data format for recording isolated lead (sink) leakage
Unit under test___________________________ Date of test:_____________________
Test Conditions
Patient Lead
System Power
Grounding/PE
RA+LA+LL
On
Closed
Off
Closed
Содержание LOGIQ E9
Страница 1: ...12 13 18 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL VERSION R6 Part Number 5573152 100 Revision Rev 2 ...
Страница 2: ......
Страница 11: ...GE DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL xi ZH CN KO ...
Страница 140: ...DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 50 Section 4 8 Site Log This page was intentionally left blank ...
Страница 753: ......
Страница 754: ......