GE
D
IRECTION
5573152-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 10 Care & Maintenance
10 - 25
10-7-10
Probe leakage current test
This test measures the current that would flow to ground from any of the probes through a patient who
is being scanned and becomes grounded by touching some other grounded surface.
NOTE: Some leakage current is expected on each probe, depending on its design. Small variations in
probe leakage currents are normal from probe to probe. Other variations will result from differences in
line voltage and test lead placement. It is abnormal if no leakage current is measured. If no leakage
current is detected, check the configuration of the test equipment.
10-7-10-1
Generic Procedure on Probe Leakage Current
The most common method of measuring probe leakage is to partly immerse the probe into a saline bath
while the probe is connected to the Ultrasound system and active. This method measures the actual
leakage current resulting from the transducer RF drive.
Measurements should be made under the test conditions specified in:
•
•
as applicable.
For each combination, the probe must be active to find the worst case condition.
DANGER
DO NOT USE THE PROBE IF THE INSULATING MATERIAL HAS BEEN PUNCTURED OR
OTHERWISE COMPROMISED. INTEGRITY OF THE INSULATION MATERIAL AND PATIENT
SAFETY CAN BE VERIFIED BY SAFETY TESTING ACCORDING TO IEC60601-1.
Figure 10-4 Set Up for Probe Leakage Current
1.
Ultrasound System
4.
Ground Switch
7.
Isolator
2.
Tester
5.
Polarity Reversing Switch
8.
Ultrasound Probe
3.
Neutral Switch
6.
Meter
9.
Saline Probe
2
3
6
5
8
7
9
1
4
A
Содержание LOGIQ E9
Страница 1: ...12 13 18 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL VERSION R6 Part Number 5573152 100 Revision Rev 2 ...
Страница 2: ......
Страница 11: ...GE DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL xi ZH CN KO ...
Страница 140: ...DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 50 Section 4 8 Site Log This page was intentionally left blank ...
Страница 753: ......
Страница 754: ......