LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 7 Diagnostics/Troubleshooting
7 - 17
7-5-3 Diagnostics Window Overview (cont’d)
Figure 7-16 Diagnostic Window Overview
The Service Diagnostics windows have some advanced features.
NOTE:
The Loop Count field must have a value of 2 or greater to activate the Advanced features.
8.) This button toggles between Standard and Advanced mode. The window below shows the
advanced features.
9.) Select the choice for Stop On Failure for the Looping and List sections:
-
Yes = the test will abort at the first failure
-
No = the test will complete even if there are failures
10.)Select the choice for Stop On Abort for the Looping and List sections:
-
List = the list will abort when you click the Abort button
-
Loop = the looping will abort when you click the Abort button
-
Diag = the diagnostic test will abort when you click the Abort button