Section 8-10 - Main Power Supply parts replacement
1.
If the 4D MC option is installed, disconnect
the USB Cable from J28 on the BEP. The
4D MC will also have to be transferred to
the replacement PS. See:
Section 8-10-4
"4D Motor Controller (4D MC)
replacement" on page 8-248
2.
Ease the PS away from the fang bracket.
Pull/slide the entire PS backwards until the
connectors to the Card Cage/Backplane
are unseated.
Lift the PS away.
Table 8-151 Main Power Supply (PS) removal