GE
D
IRECTION
5573152-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
3 - 2
Section 3-2 - Setup reminders
Section 3-2
Setup reminders
3-2-1
Average setup time
3-2-2
Setup warnings
If the unit is very cold or hot, allow the temperature of the device to stabilize before powering up. The
following table describes guidelines for reaching operational temperatures from storage or transport
temperatures. See:
2-2-1 "Console environmental requirements" on page 2-2
.
Table 3-1
Average setup time
DESCRIPTION
AVERAGE SETUP
TIME
COMMENTS
UNPACKING THE LOGIQ E9
0.5 HOUR
SET UP LOGIQ E9 WO/OPTIONS
4 HOURS
DEPENDENT ON THE CONFIGURATION
DICOM NETWORK CONFIGURATION
2 HOURS OR MORE
DEPENDENT ON THE CONFIGURATION
INSTALL INSITE / ILINK
0.5 HOUR
DANGER
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH
THE UNIT!
CAUTION
TO PREVENT ELECTRICAL SHOCK, CONNECT THE UNIT TO A PROPERLY GROUNDED
POWER OUTLET. DO NOT USE A THREE TO TWO PRONG ADAPTER. THIS DEFEATS SAFETY
GROUNDING.
CAUTION
DO NOT WEAR THE ESD WRIST STRAP WHEN YOU WORK ON LIVE CIRCUITS AND MORE
THAN 30 V PEAK IS PRESENT.
CAUTION
DO NOT OPERATE THIS UNIT UNLESS ALL BOARD COVERS AND FRAME PANELS ARE
SECURELY IN PLACE. LOGIQ E9 PERFORMANCE AND COOLING REQUIRE THIS.
NOTICE
NEVER REVERSE POLARITY ON ANY METER THAT INTERCEPTS THE POWER CORD WITH
POWER CONNECTED TO THE SYSTEM.
EVEN IN THE OFF STATE, REVERSING POLARITY ON THE POWER CAN SERIOUSLY DAMAGE
THE POWER SUPPLY.
Содержание LOGIQ E9
Страница 1: ...12 13 18 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL VERSION R6 Part Number 5573152 100 Revision Rev 2 ...
Страница 2: ......
Страница 11: ...GE DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL xi ZH CN KO ...
Страница 140: ...DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 50 Section 4 8 Site Log This page was intentionally left blank ...
Страница 753: ......
Страница 754: ......