Chapter 2 Site preparations
2 - 11
2-3-5
Networking setup requirements
2-3-5-1
Stand alone scanner (without network connection)
None.
2-3-5-2
Scanner connected to hospital’s network
Supported networks:
•
10/100/1000 Mbps Ethernet network connection
•
Up to 300 Mbps WLAN (option)
2-3-5-3
InSite Requirements
Need internet access available to be able to connect to Insite ExC.
2-3-5-4
Purpose of the DICOM network function
DICOM services provide the operator with clinically useful features for moving images and patient
information over a hospital network. Examples of DICOM services include the transfer of images to
workstations for viewing or transferring images to remote printers. As an added benefit, transferring
images in this manner frees up the on-board LCD and peripherals, enabling viewing to be done while
scanning continues. With DICOM, images can be archived, stored, and retrieved faster, easier, and at
a lower cost.
2-3-5-5
DICOM option setup requirements
To configure the LOGIQ E9 to work with other network connections, the site’s network administrator
must provide information to complete the form in
Figure 2-3 "Worksheet for DICOM Network
. Ensure that there are no spaces in any field of the form.
Entries must include:
•
A host name, local port number, AE Title, IP address and Net Mask for the LOGIQ E9.
•
The IP addresses for the default gateway and other routers at the site for ROUTING
INFORMATION.
•
The host name, IP address, port and AE Title for each device the site wants connected to the
LOGIQ E9
for DICOM APPLICATION INFORMATION. A field for the make (manufacturer) and the
revision of the device, is also included. This information may be useful for troubleshooting.
For connectivity setup information, refer to the current revision of the LOGIQ E9 Basic User Manual.
See:
Table 1-8 "Basic User Manual and Release Notes per LOGIQ E9 console" on page 1-20
Содержание LOGIQ E9
Страница 1: ...12 13 18 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL VERSION R6 Part Number 5573152 100 Revision Rev 2 ...
Страница 2: ......
Страница 11: ...GE DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL xi ZH CN KO ...
Страница 140: ...DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 50 Section 4 8 Site Log This page was intentionally left blank ...
Страница 753: ......
Страница 754: ......