GE
D
IRECTION
5573152-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 8 Replacement procedures
8 - 25
8-4-7-3
Verify and Update Vital Product Data
It is essential that Vital Product Data (VPD) is verified or updated after any hardware change to the Back
End Processor (BEP) or Front End Processors.
NOTE:
NOTE: Failure to properly program Vital Product Data could cause adverse affects on system
operation and image quality.
Refer to:
Section 5-17 "VPD Editor" on page 5-194
and
8-9-11 "VPD - GFI and MRX Configuration" on
page 8-302
for additional information.
8-4-8
Flashing the BIOS memory
1.) Boot system and insert the BIOS BLASTER disk.
2.) When prompted to load the new BIOS, press “y”.
3.) When prompted, remove the disk.
4.) Hold down the ON/OFF button to power down the system.
5.) Reboot the system, and look for the new BIOS number.
8-4-9
Setup after Software loading
NOTE:
Use this procedure if all partitions on the HDD were erased during the software load.
NOTE:
If only the software on C:\ has been replaced or updated, go to: 8-4-10 "Verifications after the
software has been loaded" on page 8-26.
1.) Restore the Patient Archive and System Configurations from the backup you made before the
software load. For instructions, see “Data Backup and Restore” in the User Manual/User Guide.
2.) With your recordings from before the software load available, continue with the setup instructions
starting in:
Section 3-7 "Configuration" on page 3-31
. Correct the settings as needed.
3.) Continue with:
Section 3-9 "Connectivity Setup and Tips" on page 3-35
. Based on your recordings,
correct the settings as needed.
4.) Continue with:
Section 3-10 "Setup paperwork" on page 3-36
. Based on your recordings, correct
the settings as needed.
5.) After a software load, you should always calibrate the Front End, as described in:
Offset Calibration" on page 6-18
.
Содержание LOGIQ E9
Страница 1: ...12 13 18 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL VERSION R6 Part Number 5573152 100 Revision Rev 2 ...
Страница 2: ......
Страница 11: ...GE DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL xi ZH CN KO ...
Страница 140: ...DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 50 Section 4 8 Site Log This page was intentionally left blank ...
Страница 753: ......
Страница 754: ......