LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 8 Replacement procedures
8 - 111
8-6-3-2
LCD Monitor V2 Arm assembly (Daeil) replacement (used in R4.x production)
Table 8-72 LCD Monitor V2 Arm assembly (Daeil) removal - R4 production
Steps
Corresponding Graphic
1.
Remove the LCD Monitor.
2.
WARNING
WARNING
WHEN REMOVING THE LCD ARM, KEEP
THE ARM LOCKED, AS SHOWN TO
ENSURE THE LCD ARM ASSEMBLY IS
IN THE LOCKED POSITION UNTILL THE
ARM AND LCD ARE INSTALLED. THE
SPRINGS TO SUPPORT THE LCD CAN
CAUSE THE ARM TO SPRING OPEN
CAUSING SEVERE PERSONAL INJURY
AND PROPERTY DAMAGE.
3.
Remove the Bulkhead Cover.
4.
Disconnect the LCD cables; video (
P10
)
and the power (
P3
) from the LCD.
The Video Cable has a retainer and screw.
Loosen the screw with a small, flat blade
screwdriver.