GE
D
IRECTION
5573152-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 8 Replacement procedures
8 - 17
8-4-6-5 Loading the Application Software (cont’d)
6.) Remove the applications disk from the DVD drive.
7.) From the Start Applications window, select “
Start
” (if the CD is still in the tray; otherwise, the system
starts up automatically).
The System powers down and reboots.
The first time the application loads after a SW reload, you may see: the error shown in
.
Select
OK
and reboot the system.
During the reboot process, the system
may
prompt for the Basic Option string (valid license) (see:
If the Basic Option string is not saved in the Option Dongle, or if this is a software upgrade, the SW
License window may open.
NOTE:
If only the title bar is visible in the lower left corner of the monitor, double-click the SW License
title bar to expand the window.
NOTE:
All systems must have at least the Basic Option string for the software to run. Any single Option
String may enable multiple options (i.e. Basic).
Demo Option strings turn on all options for a limited period of time. Do Not leave a demo Option
string in a unit when it is sold to a customer.
NOTE:
If the system does not display an image or a good image when you boot up, Do Not troubleshoot
it as a problem yet. Continue with this install.
8.) Update the Vital Product Data (VPD) information.
Proper Vital Product Data coding is essential to proper system function and optimization. For
information on updating VPD, see:
8-4-7-3 "Verify and Update Vital Product Data" on page 8-25
Figure 8-6 SW Reload Error Message
Figure 8-7 Enter Option String
Содержание LOGIQ E9
Страница 1: ...12 13 18 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL VERSION R6 Part Number 5573152 100 Revision Rev 2 ...
Страница 2: ......
Страница 11: ...GE DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL xi ZH CN KO ...
Страница 140: ...DIRECTION 5573152 100 REV 2 LOGIQ E9 SERVICE MANUAL 4 50 Section 4 8 Site Log This page was intentionally left blank ...
Страница 753: ......
Страница 754: ......